Worldwide supplier of underwater optical sensors since 1994
sales: (603) 642-4921  ▪  support: (603) 775-7707  ▪  email: seapoint@seapoint.com
seapoint navigation bar
 

Seapoint Fluorescein Fluorometer
Download pdf Datasheet or User Manual

  FEATURES
• Very low power requirements
Can be used in pumped or open deployments
Small size
6000 m depth capability
Good ambient light rejection
Linear output with Fluorescein concentration
Four programmable ranges
Low temperature coefficient
Low offset voltage does not require adjustment
Pin compatible with other Seapoint Sensors
Interfaces easily with data acquisition systems
Rugged, corrosion-free materials

APPLICATIONS
Profiling, Moored, Towed, or In-Line Measurements
Dye Trace Studies
Mixing and Dispersion Studies
Leak and Spill Detection
Flow Measurement
Reverse Osmosis Recovery Measurement

Seapoint Fluorescein Fluorometer Top View
 

Seapoint Fluorescein Fluorometer


DESCRIPTION
The Seapoint Fluorescein Fluorometer (SFF) is a high-performance, low power instrument for in situ measurements of Fluorescein. Its small size, very low power consumption, high sensitivity, wide dynamic range, 6000 meter depth capability, and open or pump-through sample volume options provide the power and flexibility to measure Fluorescein in a wide variety of conditions. The SFF uses modulated green LED lamps and a narrow-band excitation filter to excite Fluorescein. The fluorescent light emitted by the Fluorescein passes through an orange emission filter and is detected by a silicon photodiode. The low level signal is then processed using synchronous demodulation circuitry which generates an output voltage proportional to Fluorescein concentration. The SFF may be operated with or without a pump. The sensing volume may be left open to the surrounding water, or, with the use of the supplied cap, can have water pumped through it. Two control lines allow the user to set the range to one of four options. These lines may be hardwired or microprocessor controlled to provide a suitable range and resolution for a given application. The sensor is easily interfaced with data acquisition packages; a 5 ft. pigtail is supplied. Custom configurations are available.  

 
 
SPECIFICATIONS

• Power Requirements:

• Output
• Output Time Constant
Power-up Transient Period
Excitation Wavelength
Emission Wavelength
Sensing Volume
Minimum Detectable Level
 
Sensitivity/Range



 

• Temperature Coefficient
• Depth Capability
• Weight (dry)
• Operating Temperature
• Material
• Underwater Connector


8-20 VDC, 15mA avg., 27mA pk.
0-5.0 VDC
0.1 sec.
< 1 sec.
475
nm CWL, 30 nm FWHM
530 nm CWL, 30 nm FWHM
340 mm3

0.02 µg/l
 
Gain
30x
10x
3x
1x
Sensitivity, V/µg/l
1.0
0.33
0.1
0.033
Range, µg/l
5
15
50
150

< 0.3%/°C
6000 m (19,685 ft)
1000 g (2.2 lbs)
0°C to 65°C (32°F to 149°F)
Rigid polyurethane
Impulse AG-306/206 (others available on request)

 
DIMENSIONS
Schematic of Seapoint Fluorescein Fluorometer   

Home | Products | Global Sales | Download | Contact Us | Site Map